A.1.4 Baseline Noise & Distortion Test - Histogram Analysis
Connect the evaluation board’s analog input(s) to circuit ground.
Select the Histogram window by clicking on the Histogram tab at the top of the screen.
Click on the Acquire button. Within the Acquire sub-panel, set Select Number of Samples to Capture to
524288. Set the Channel(s) to Capture to the desired channel(s).